Product Details

  • Standard LFE calibration accuracy is +/- 0.72% of Reading (not % of Full Scale)
    traceable to the National Institute of Standards and Technology (NIST)
  • Rigid construction and lack of any moving parts make the LFE calibration stable.
  • Only physical damage or particulate deposition in the element will cause a calibration shift.
  • Standard Accuracy: +/-0.72% to +/- 0.86% of Reading
  • Repeatability: 0.1%
  • Response time: 16ms
  • Turndown: 20:1
  • Achievable with accurate DP device